How to check, clean, store and calibrate a force gauge, when to send it for repair, and what Australian standards say about keeping force measurements reliable.
A force gauge stays accurate when you check it at zero before use, keep it clean and dry, protect it from overload and drops, store it in its case, and have it calibrated periodically by a NATA-accredited laboratory. Most faults we see trace back to an overload or a knock, not to normal wear. This guide covers the routine, the warning signs and the calibration decisions, so your results hold up in an audit.
A force gauge is a small instrument that makes big decisions. It signs off the pull strength of a weld, the closing force of a pack, the peel of a seal or the insertion force of a component. If the gauge drifts, product passes that should fail, or good product gets scrapped. Either way, you pay for it in rework, complaints or findings at audit.
Maintenance is separate from calibration. Maintenance is what your team does every day and every week to keep the gauge in good condition. Calibration is the periodic comparison against traceable reference standards that tells you how far the reading is from true. You need both. Good maintenance makes calibration results stable from one cycle to the next, and calibration confirms that your maintenance is working.
The Institute of Measurement and Control (InstMC) suggests that confidence checks for drift, noise, non-linearity and hysteresis belong in a regular maintenance program. That is a useful way to frame the work: small, scheduled checks between calibrations, recorded so you can see a trend.

The quickest check is also the most valuable. With nothing attached and no load applied, switch the gauge on and look at the zero reading. The InstMC force measurement guide advises users to monitor the output at zero force as a regular check on whether the transducer has been overloaded. If the zero reading has moved from what you normally see, stop and investigate before you test anything.
If your quality system allows, add a periodic check with a known reference mass or a verified reference gauge, at a force close to what you normally measure. Record the result against a limit you have set. This is not a calibration, and it does not replace one. It tells you whether the gauge is still behaving between accredited calibrations. We cover the same logic for weighing instruments in our daily checks for audit-ready balances.
Cleaning looks harmless, and it is where well-meaning staff do damage. The principle is gentle and dry.
Food and pharmaceutical sites often have sanitising routines that rely on alcohol wipes. Check what your gauge’s manufacturer permits before applying them. If the manual forbids alcohol, clean the fixtures and contact surfaces instead, and keep the gauge body to a dry wipe. Where the manual is silent or you are unsure, ask us, because a cleaning step that suits your hygiene procedure but not your instrument is a quiet source of damage.
The sensor inside a force gauge is a small, precise element. Think of drops and overloads as the main hazards, with heat, moisture and contamination close behind.
The Shimpo manual says to keep the gauge away from water, oil, dust, chemicals and temperatures above 40°C, and to return it to its case when not in use to prevent accidental-drop damage. In a New South Wales plant, that matters in summer: a gauge left on a bench near a hot machine, or in a vehicle, can easily exceed that limit.

For long storage, Safe Work Australia’s code says plant should be stored to minimise damage such as corrosion and, after extended storage, recommissioned with the same inspection and testing used at first commissioning. For a gauge, the practical version is simple: after it has sat unused for months, do your zero check, a reference check and, if it matters for your results, send it for calibration before putting it back into service.
Take a gauge out of use and tag it when you see any of the following.
| Sign | What it can mean | What to do |
|---|---|---|
| Overload indication persists with no load applied | Nidec’s troubleshooting table says the internal load cell may be broken, commonly after a drop or overload | Remove from service and send for repair |
| Display does not return to normal after an overload is removed | Shimpo’s table says the gauge should be sent for repair | Remove from service and send for repair |
| Persistent EEPROM error codes | Shimpo lists these as requiring repair | Do not keep using it. Send for repair |
| Persistent low-battery, display or charging faults | Nidec says these also warrant repair | Replace the battery once, and if the fault returns, send for repair |
| Zero reading has shifted from your logged norm | Possible overload or damage, per the InstMC guidance on zero-force monitoring | Investigate, then calibrate or repair before further use |
Two habits make the biggest difference. First, keep a short incident log: any drop, overload warning or unusual reading goes in with the date and the operator. Second, after any such event, do not simply carry on. Check zero, run a reference check, and if anything is off, quarantine the gauge. Results produced since the last good check may need review, and your quality team will want to know that quickly.
There is no single number that suits every gauge. The Australian Standard on force-measuring systems, AS 2193:2005, says systems should be maintained in good working order, serviced according to frequency and conditions of use, and calibrated periodically so results stay reliable and consistent. NATA also publishes calibration interval guidance by equipment type that refers to AS 2193 for force testing machines.
In practice, the interval you set should reflect:
Manufacturers that run a formal quality assurance program tend to tie these intervals to documented risk, and that is a sound approach: write down why you chose the interval, review it after each calibration, and keep the certificate with the gauge’s record.
For force calibration in Australia, the National Measurement Institute provides services for transducers, load cells and proving rings in accordance with AS 2193, ISO 376 and ASTM E74. When you send a gauge to a commercial laboratory, ask for a NATA-endorsed certificate with traceability to national standards, so an auditor can follow the chain. Our page on traceable calibration explains what that traceability looks like, and our NATA calibration page covers what accreditation means for your records.
If you work in a regulated sector, your calibration records face scrutiny. Our guides to GMP calibration requirements, AS9100 calibration and ISO 13485 calibration set out what quality managers typically need to show.
Most force gauge problems we see did not start with age. They started with one bad moment: a gauge dropped from a bench, a load that went past capacity, or a wipe-down with the wrong cleaner. The instrument then kept running, kept producing numbers, and nobody questioned them until the next calibration.
That is why we push clients towards a simple discipline. Check zero. Log the result. Treat every overload or drop as an event. Then use calibration as a confirmation of your routine, not as the first time anyone looks at the gauge.
PCS Precision is NATA accredited and ISO 9001 certified, and we support multiple equipment brands regardless of where you bought the instrument. We offer on-site and workshop calibration, preventative maintenance and repairs, with 24/7 emergency support for urgent faults. For a quality manager, that means one partner who can calibrate the gauge, tell you what the as-found result means, and fix it if it needs fixing.
Proactive care reduces unplanned downtime. A gauge that fails on the day of a customer audit or a batch release is expensive. A gauge that was checked, logged and calibrated on a schedule is not.
Kaz
Booking is quick when you give the laboratory the right information up front. It saves back-and-forth and lets us quote accurately. Have these details ready.
On-site work suits large setups and gauges mounted on production equipment, while workshop calibration suits portable units that can be sent in. Our article on on-site vs in-lab calibration walks through how to choose. If cost is a concern, see why NATA quotes cost more, which explains what you are paying for.
If your gauge is due, has been dropped or is showing a fault, we can help. We provide NATA calibration, traceable calibration, preventative maintenance, repairs and equipment hire, and we work across multiple brands. Book a calibration or request a quote and tell us the gauge details listed above. If you need a replacement or an additional instrument, browse our measuring equipment range.
Check the zero reading before each session and compare it with your logged norm. A shift can indicate the transducer has been overloaded, as the InstMC guide notes. Add a periodic reference check at a force close to your working range, and record the results so you can see drift early.
Check your manufacturer’s manual first. Nidec’s manual says not to use volatile chemicals such as alcohol. A dry soft cloth is the safe default, and a cloth dampened with diluted detergent and wrung out is the next step. Switch the gauge off and avoid pressing on the sensing element.
AS 2193 calls for periodic calibration and servicing based on frequency and conditions of use. Heavy use, near-capacity loads, harsh environments and overload events all argue for shorter intervals. Review the interval after each calibration using the as-found result.
Stop using it. Nidec’s troubleshooting guidance says a persistent overload indication with no load may mean the internal load cell is broken, often after a drop or overload. Tag it out of service and send it for repair, then review results produced since the last good check.
We advise against it. Manuals such as the HBO HF series warn against disassembling the gauge yourself. Opening the case can disturb the sensor and void the instrument’s traceability, so use a qualified service provider and recalibrate after repair.